Introduction
The lumbar spine experiences compressive preloads up to 800N during activities of daily living (ADL). However, kinematics of lumbar disc prostheses under large preloads have not been reported. In our experience, disc prostheses with articulating bearings tend to “bind” under large preloads, resulting in degradation of motion quantity and quality.
Purpose
To test the hypothesis that quantity and quality of motion of lumbar segments implanted with compressible non-articulating disc prostheses will not be significantly affected by compressive preload magnitude.
![]() |
| Figure 1: Schematic of a lumbar spine mounted in the biomechanical test set-up (left).A follower load is applied through the centre of rotation of each segment via the Preload Cable. A moment is applied and the resulting ROM is measured with the attached sensors (right). |
Methods
Eight human cadaveric lumbar spines (L1-S1, 44±6.5 yr) were tested in flexion (8Nm) and extension (6Nm) under 0N, 400N and 80